Contact an Account ManagerTriazine herbicides from water
Summary
Type: SPE
Application #: 303240
Catalog#: 730108 - SPE Cartridges, Chromabond, HR-P, 3ml, PP, 200mg 30/pk
Detection: GC
Conditions: Conditioning: Suck 1 column volume Methanol , then 1 column volume dest. water through the coliumn.
Attetion: The column must not run dry!
Sample application: Suck 1000 ml water sample through the column.
Drying: Nitrogen stream
Elution: With 3 x 1 ml Methanol/Aceton (3:2, v/v) and then evaporate to the needed volume.
Authors: Macherey-Nagel
Source: SPE Department
Details
Legend: Substances Recovery rates (%)
Atrazin 98
Ametryn 94
Desisopropylatrazin 99
Desethylatrazin 98
Terbuthylazindesethyl